top of page

ValProbe RT Temperature Data Loggers - Bendable Sensors

Product Tags

The ValProbe RT Temperature Data Loggers with bendable sensors provide a vast temperature range from -85°C to + 400°C and are available as single and dual and five channel bendable sensor.

Request A Quote

State
Description

ValProbe RT Temperature Data Loggers with Bendable Sensors - Single, Dula and 5-channel bendable logger


The ValProbe RT Temperature Data Loggers with bendable sensors provide a vast temperature range from -85°C to + 400°C. They are available as single and dual and five channel bendable sensor.


RTD technology delivers unrivalled measurement accuracy and the new logger electronic design guarantees long battery life. Operator programmable sample rates starting at 1 second, allow collection of up to 100 000 data points per sensor.


The antenna diversity concept of the temperature data loggers secures an unmatched RF communication reliability.

Key Features

1 and 2-Channel Bendable Logger

  • Temperature range for logger sensor: -85°C to 400°C

  • Single and dual bendable sensor available

  • Sensor length 12, 24, 36”

  • Sensor diameter 2.4mm; 0.095“


5-Channel Bendable Logger

  • Temperature range for logger sensor: 0°C to 400°C

  • Five bendable sensors

  • Sensor length 24”

  • Sensor diameter 2mm; 0.078“

Applications
  • Steam Sterilizer

  • Dry Heat Sterilizer

  • Steam in Place (SIP)

  • Water Cascade/Fall Sterilizer

  • Incubators

  • Stability Chambers

  • Freezers

  • Freeze Dryer/Lyophilization

  • Vessels

Specifications

Specifications

1 and 2-Channel Bendable 

 5-Channel Bendable

Sensor Type 

Single and Dual Sensors 

 Five Sensors

Sensor Length

12, 24, 36" inches305, 610, 915 mm

24" inches610 mm

Tip Diameter

0.095"; 2.4mm

0.078"; 2mm

Measurement Range

-85°C to 400°C

0°C to 400°C

Accuracy

-85°C to 140°C,    ±0.1°C140°C to 400°C    ±0.25°C

0°C to 140°C,    ±0.1°C140°C to 400°C    ±0.25°C




Kaye ValProbe RT Generic Specifications

Logger Dimensions 

Hight: 1,9'' / Diameter 1,4''


48mm / 36mm

Logger Material

Stainless Steel 316L and Peek 

Battery

Field replaceable - 3.6V Lithium

Sampling Rate

Starting at 1 sec

Data Storage

100.000 Samples


 retained in non-volatile memory

Real-Time Clock Accuracy

< 15sec/day

 


Kaye ValProbe RT Generic Specifications

Calibration

NVLAP (NVLAP LAB CODE 200913-0)/DAkkS Calibration

Verification

Automated User


 Verification capability

Sensing Element

Precision Platinum RTD

Environmental Temperature

-85°C to 140°C

Environmental Pressure

0-5bar absolute

Environmental Humidity

0-100% condensing

Regulatory Compliance

UL and CE



Enquire Now

Fill out the form below and together we will find a perfect customised solution tailored for your business.

State

© 2021 CISCAL. All rights reserved.

bottom of page