Hyperlab wine analyzer represents the most advanced solution for analytical control automation in the oenological field. 

Enquire Now

State

Description

Hyperlab wine analyzer represents the most advanced solution for analytical control automation in the oenological field. Over 500 installations and 15 years of quality and reliability in the best enological laboratories.


The results of the analyses are extremely accurate and permit the user to work in an extremely time-efficient manner thus allowing for both quality analysis improvement and cost-benefit optimization.


Hyperlab is an automatic multi-parametric analyzer with Flexible Random Access, i.e. it can fully perform all the manual operating procedures automatically.


The sampling arm, controlled by a microprocessor, takes the wine sample, dilutes it (if necessary), and adds the required reagents. The sample and reagents are mixed in a reaction cuvette, where they remain at a constant temperature until the reaction is complete.


The result of the reaction is read at a specific wavelength, and the recorded absorbance is converted into concentration using calculation algorithms.

Benefits

BEST QUALITY AND ECONOMICAL

Hyperlab automation guarantees the greatest data results accuracy and precision, thus significantly reducing analysis costs by about seven times when compared with manual methodologies.


With Hyperlab reagents are reduced to minimum: a single determination requires less than 300 microlitres.

METHODS IN MEMORY

Acetic acid, Citric acid, D-Gluconic acid, D-Lactic acid, L-Lactic acid, L-Malic acid, Pyruvic acid, Tartaric acid, Acetaldehyde, Anthocyanin, Alpha Amino nitrogen, Ammoniacal nitrogen, Calcium, Catechins, Chlorides, Iron, Glycerol, Glucose/Fructose, Glucose/Fructose (for high sugar concentration analysis on musts), Magnesium, Total polyphenols, Potassium, Copper, Free and Total SO2, Optical reference solution.


Other methods:

IC (colour intensity 420+520), shade (420/520), ICM (modified colour intensity 420+520+620), GRB (% yellow, red and blue colour on total colour. This is an indispensable determination for products undergoing refining or micro-oxygenation process).The optical reference solution gives the technical specifications of the instrument regardless its kit performances.The best way to demonstrate Hyperlab precision is the use of bicromatic optical solution (suggested by IUPAC - International Union of Pure and Applied Chemistry).

CALIBRATIONS

Reagent blank subtraction, 1 to 8 calibrators for single test. Linear and non-linear regressions with three types of data extrapolation: cubic, polynomial and log-logit.

INNOVATIVE SOFTWARE AND HARDWARE

Innovative software: Windows system, user-friendly functions, touch screen technology, switch-on/off with programmable start-up, on-line support.


Monitor: Information on programmed methods, selected analyses, n° of analyses to be run, calibration and check status.


Reagent panel: Displays the samples position and their volumes.


Samples: Samples status display. Possibility of adding, removing or modifying samples during the analysis.


Working lists: Unlimited working lists can be used simultaneously. Tests may be added or removed and automatically repeated if wrong. Inspection of kinetics reaction can be performed for single tests.

DATA PROCESSING AND TESTS

Data processing

Results can be re-calculated at the end of the analysis by correlating them to one or more samples with fixed value. The new regression line will show the value obtained by the re-calculation.


Tests


All the analyses regarding a spectrophotometric determination can be automatized. 

Key Features

  • VOLUMES Sample: 2-300µl, precision 1% at 10µlReagent: 5-350 µl, precision < 1% at 250µlReaction (sample + reagent): 220-350µlMixing in cuvette after dispensation


  • DILUTIONAutomatic programmable dilution for each sampleFactors: 1:1, 1:4, 1:10, 1:40, 1:100


  • ANALYTICAL SPEED125...200 analyses/hour - Hyperlab Plus100...180 analyses/hour - Hyperlab Basic


  • TEMPERATURE CONTROL Refrigerated reagent at 15°C below ambient temperatureReading cuvette with programmable heatingReagent preheating in sucking needle


  • TYPES OF TESTS Endpoint, Bichromatic endpoint, Differential endpoint, Differential endpoint sample blank, Derivatives, Fixed time, Kinetic.


  • SAMPLES TRAY: 60 samples positions


  • REAGENTS TRAY: 30 positions refrigerated tray


  • SAMPLING NEEDLE: Glass ceramics stroke needle, capacitance sensor, reagent pre-heating 


  • REACTION CUVETTE ROTOR: 80 BIONEX® cuvettes replaceable singularly, over 20.000 tests


  • WASHING STATION: 6 dispensing needles for washing solution, 6 sucking needles, 1 drying needle, 8 washing cycles for each cuvette


  • OPTICAL GROUP1 halogen lamp (6V, 10W) with extended UV emission, 2 focusing lenses, 10-position interference filters, 8 positions provided with interference filters of 340, 420, 520, 546, 578, 620, 650, 700 mm, 1 free position and 1 for dark reading.


  • DATA RE-PROCESSING Results can be re-calculated at the end of the analysis by correlating them to one or more samples with fixed value


  • INNOVATIVE SOFTWARE Extremely user-friendly. Optimized random system allows the user to work efficiently and comfortably. Operating system: Windows2x20lt plastic tanks for washing and discharge water- Hyperlab Plus2x5lt plastic tanks for washing and discharge water- Hyperlab Basic


  • SAFETY SHIELD SENSOR on request


  • DIMENSIONS: LxDxH 60x66x50 cm


  • WEIGHT: 38Kg 


Enquire Now