top of page

Laser Calibration: When and How to Do It

Use a risk-based interval typical practice: 6 to 12 months for regulated work; shorter if critical, high-use, or harsh environments. A fixed number isn’t mandated by ISO/IEC 17025, intervals must be justified and records kept.


What triggers a calibration? On installation/commissioning, after any impact/repair, after major software/firmware changes, when drift is detected, and at your defined interval.

Audit-ready results in Australia: Use labs with NATA-endorsed certificates showing SI traceability via Australia’s National Measurement Institute (NMI), with measurement uncertainty reported.




What is Laser Calibration? 

Laser calibration is a comparison of your instrument’s readings against a more accurate, traceable reference to quantify error and report expanded uncertainty (95% confidence). In Australia, NATA requires metrological traceability to SI units, typically through NMI, and ISO/IEC 17025 sets the competence framework labs are assessed against.


Common Categories:

  • Dimensional: Laser interferometry for machine tools/CMM axes; generates compensation tables to correct positioning errors.

  • Radiometric: Laser power/energy meters verified against NMI-traceable standards; checks responsivity and linearity.

  • Spectral: Wavelength checks of lasers/wavelength meters against stabilised references or transfer standards; uncertainty stated in nanometres per the lab’s scope.

  • Beam diagnostics: Beam profile/divergence/M² checks to ensure process or research performance matches spec.

  • Construction lasers (levels): Practical level/line checks and, if out, full lab calibration.


Compliance in Australia 

  • NATA & ISO/IEC 17025: NATA accredits labs to ISO/IEC 17025, providing independent assurance that methods, uncertainty, and traceability are sound. NATA-endorsed certificates are widely recognised, including via ILAC.

  • Traceability & uncertainty: NATA’s Metrological Traceability Policy explains how results must be linked to national standards (commonly NMI) and how uncertainty is established and reported. 

  • Laser safety labelling/classification: Follow ARPANSA guidance and AS/NZS IEC 60825 series (equipment classification, user guidance).

  • Workplace controls (construction): Safe Work Australia states Class 3B and 4 lasers must not be used for construction work. Use Class 1/1M/1C/2/2M/3R only.

  • Sector drivers: TGA adopts PIC/S GMP for medicines (calibrated, traceable instruments and records); FSANZ requires at least one thermometer accurate to ±1 °C in food businesses (handy for instrument verification in HACCP).


When to Calibrate: By Risk & Use Case 

Set intervals with evidence. Consider safety/quality risk, usage hours, environment (heat, vibration), historical drift, firmware changes, and audit expectations. Document the rationale in your SOP.

Application

Typical triggers

Suggested interval (guide only)

Standard/driver

Machine tools / CMM axes

Commissioning, after crash or ball-screw work; tolerance changes

6–12 months for production; shorter if tight tolerances

ISO/IEC 17025 conformity; OEM specs; NATA traceability; laser interferometer methods per vendor guidance. 

Laser power/energy meters

Before validation/R&D campaigns; after sensor replacement/impact

6–12 months; verify at operating wavelengths and expected ranges

NATA traceability via NMI optical services; lab scopes list ranges/uncertainties.

Wavelength meters/spectrometers

Before critical experiments; after firmware/hardware change

≈12 months for regulated labs; risk-based in research

NMI optical standards; NATA-endorsed certificates show SI traceability and uncertainty. 

Construction laser levels

After drops/shock; if site check fails

Site check monthly; lab calibration as per contract/spec

Field check per RedBack method; if out, book NATA calibration.

How to Calibrate: Procedures and Checklists

ree

A. Laser Interferometry (Machine Positioning)

What you’re doing: Using a laser interferometer (or tracker with interferometry) to measure linear errors, backlash, straightness, squareness, pitch/yaw/roll and then generating axis compensation tables in the controller.


Set-up essentials (checklist):

  • Stable environment (temp, air flow); warm-up machine and optics.

  • Align optical path; use a retroreflector/SMR or plane mirror targets.

  • Log environmentals (air temp/pressure/humidity) for refractive index compensation.

  • Verify laser reference status and traceability; check beam quality.


Run-through (summary):

  1. Baseline sweep on each axis (up/down) for linear error and reversal.

  2. Cross-tests for straightness and squareness.

  3. Rotary/axis tests if applicable.

  4. Upload compensation tables; re-run for as-left verification; issue uncertainty-backed report.

Many Australian shops use systems like Renishaw XL-80 or API trackers; both depend on interferometry with traceable wavelength standards.


B. Laser power/energy meters

Aim: Compare DUT readings to a NATA-traceable reference at relevant wavelengths and power/energy levels; check linearity and responsivity; report expanded uncertainty (k≈2). Use NMI-traceable standards or transfer artefacts. 

Steps (bench):

  1. Inspect sensor head; confirm damage/contamination-free.

  2. Stabilise source; set wavelength compensation.

  3. Apply points across the working range (up/down); hold steady; record as-found.

  4. If allowed, adjust cal factors; repeat for as-left; capture ambient conditions and drift notes.

  5. Include traceability and uncertainty budget on the certificate.


C. Wavelength (Lasers/Wavelength Meters)

Aim: Validate wavelength accuracy against stabilised references (e.g., iodine-stabilised He-Ne or frequency-comb-derived transfer standards) or accredited transfer standards; verify across your working range; report uncertainty in nm. Use a lab with appropriate scope. 

Steps:

  1. Warm-up the DUT; set to nominal lines (e.g., 632.8 nm).

  2. Compare to reference; note offsets; repeat across range.

  3. Report as-found/as-left, stability, and uncertainty with full traceability chain.


D. Field Check for Construction Laser Levels (Quick Site Method)

Use when you need a fast go/no-go on site.


5-step check (horizontal line):

  1. Set the laser ~10 m from a wall; mark the beam.

  2. Rotate 90°; mark again; repeat for 180° and 270°.

  3. All marks should align within the maker’s tolerance.

  4. If out, don’t “tweak” in the field, book a NATA calibration.

  5. After knocks/drops, re-check before use.


Documentation Auditors Expect 

Have these items on every certificate/SOP checklist:

  • NATA-endorsed certificate and scope reference (ranges and CMCs).

  • SI traceability statement (chain to NMI or another NMI via ILAC).

  • Method (interferometry, radiometry, spectral), as-found/as-left data, and environmental conditions.

  • Expanded uncertainty (coverage factor) and the decision rule used.

  • Technician and reviewer sign-off; due date/next interval; digital record retention.


Safety & legal obligations in AU 

  • Laser safety classes: Follow AS/NZS IEC 60825 classification and ARPANSA guidance. Label devices with class, power, wavelength, use signs, and implement controls per class.

  • Construction work: Do not use Class 3B or 4 lasers for construction tasks; they present significant eye/skin hazards and require strict controls.

  • Training: Consider Laser Safety Officer/Supervisor training and consult your state/territory regulator for local requirements.


Sector Call-outs 

  • Pharma/biotech: The TGA adopts PIC/S GMP; keep periodicity risk-based and show it in your validation/CAPA trail. Reference NATA-endorsed calibration in your VMP/SOPs.

  • Food & beverage: FSANZ requires at least one probe thermometer accurate to ±1 °C; if you use IR “laser” thermometers for checks, validate against a probe and document.

  • Research & engineering labs: Mixed dimensional/spectral/power work—ensure the lab’s scope actually covers your range and uncertainty needs. NMI optical and length services are the national reference.


Choosing a Provider 

Quick checklist:

  • NATA-accredited for the optical/laser scope you need (check the lab’s Scope of Accreditation).

  • Traceability to NMI stated on certificates.

  • Fit-for-purpose uncertainty at your wavelength/power/range.

  • On-site vs lab capability (e.g., on-site interferometry; lab-grade radiometry).

  • Turnaround & logistics that suit validation windows.

  • Digital certificates/asset portal for audits.


Common Drift Causes & Troubleshooting 

  • Heat and air turbulence shifting interferometer paths, control HVAC, allow warm-up.

  • Vibration and transport shock, use isolation mounts; re-check after moves/impacts.

  • Optics contamination, clean lenses/windows per OEM.

  • Fibre connector wear, inspect ferrules; replace worn leads.

  • Detector ageing (power meters), trend responsivity over time; adjust intervals if drift grows.

  • Firmware changes, treated as a calibration trigger with as-found/as-left records.


Glossary 

  • Traceability: An unbroken chain of comparisons to standards, with stated uncertainties, up to SI units (usually via NMI in Australia). 

  • Expanded uncertainty (95% CL): Reported uncertainty multiplied by a coverage factor, often k≈2, giving ~95% confidence.

  • Responsivity: Ratio of detector output to incident optical power (e.g., V/W).

  • Linearity: How constant responsibility is across the operating range

  • Beam profile: Intensity distribution across the beam cross-section.

  • Compensation table: Controller file that corrects axis errors at positions.

  • MPE: Maximum Permissible Exposure, safety concept defined in the laser standards/guides.


How CISCAL Helps 

  • NATA-accredited, ISO/IEC 17025 calibration for laser interferometers, laser power/energy meters, wavelength meters/spectrometers, construction laser levels, and optical instruments.

  • Nationwide support (NSW, VIC, QLD, WA, SA, TAS, NT), onsite and lab options.

  • Advanced optical tools and SI traceability via NMI; digital certificates with uncertainty and decision rules.

  • Fast turnaround aligned to qualification/validation windows.

FAQs


© 2021 CISCAL. All rights reserved.

bottom of page